8.10.
LINEARITY LIMIT PROGRAMMING
The maximum measuring limit is given in the reagent technical sheet. If result is beyond this limit, it is
recommended to dilute the sample or rerun the test.
8.11.
NORMAL HIGH AND LOW PROGRAMMING
Normal High & Low values border the acceptable pathological range and they are noticed in the
reagent technical sheet. It is also called the usual expected values. Flags will be shown on abnormal
result.
8.12.
QUALITY CONTROL FREQUENCY PROGRAMMING
At the end of programming, select the frequency of analysis quality control checking : whether daily or
periodically. A signal will remind you to do the quality control every day or once every x days. Result of
the QC can be found at « CONFIGURATION », « QUALITY CONTROL » and « DISPLAY ».
SAVE the method which has been programmed