BASIC
- 42 -
8.14.
MULTI-STANDARDS CALIBRATION
“ESTRADIOL” analysis includes several standards. It is necessary to indicate if it should be
recalibrated or if the calibration can be preserved. So choose with the keys “Repeat? ” or “OK”.
If the user press “Repeat?”, he will see :
It is then possible to modify the concentration values of each standard by pressing on the
corresponding function keys (of F1 with F6) then while entering the new value using the numeric
keypad. The “VALID” key makes it possible to reach the following value.
If the values on display are correct, validate. Otherwise modify using numerical keys, then validate.
The measure starts with reagent blank reading. This will be deducted automatically from other
readings :
If the user doesn’t wish to measure the reagent blank, he has to press “Esc” key.