SYSTEM DESCRIPTION 3
31
8026364 / V1-0/2022-03|SICK
SUPPLEMENTARY OPERATING INSTRUCTIONS | VMS4200/5200
Fig. 18: Dynamic scale value switchover (length dimension as switchover point)
Legend
1 Switchover point 2,000 mm
2 Identification label for scale interval value 1
3 Identification label for scale interval value 2
In example 1, the multi-dimensional measurement system is working with the two scale
interval values 5 mm x 5 mm x 5 mm and 10 mm x 5 mm x 5 mm. An object length of
2,000 mm has been defined as the switchover point during commissioning.
Objects with a length less than or equal to 2,000 mm are measured with scale interval
value 1 (5 mm x 5 mm x 5 mm). For objects with a length greater than 2,000 mm, the multi-
dimensional measurement system automatically uses scale interval value 2 (here 10 mm)
for the length dimension due to the detected object length, and continues to use scale
interval value 1 (here 5 mm x 5 mm) for the width and height.
Example 1