Centrifuge Module Interface Operating Instructions Introduction
092D0001-02 Rev. A page 21
Test List Screen
Note Your logon and password authorization might not allow access to
this screen.
Note The test list includes only those tests defined as Cap On tests.
The Centrifuge Test List screen displays tests selected as pre-tests for the
Centrifuge module. Refer to the ADVIA Automation Solutions Operator’s
Guide for general information on the Test List screen.
Note To prevent the system from routing samples to the Centrifuge
module, transition the Centrifuge module to the Shutdown state. Do not
use the Test List screen to prevent samples from routing to the Centrifuge
module. Clearing the Centrifuge Pretest check box under the Enabled or
Selected columns can cause spun samples to return to the Centrifuge
Module and then be routed to the SIQ gate or tray.
Figure 12: Centrifuge Station Test List Screen