Introduction Centrifuge Module Interface Operating Instructions
page 22 092D0001-02 Rev. A
Diagnostics Screen
The Centrifuge Diagnostics screen displays interface status information,
interface gate information, and a log of interface activities.
Figure 13: Centrifuge Station Diagnostics Screen
Diagnostics Screen Areas
Note The buttons and checkboxes are only user-accessible as described in
the following section.
The Diagnostics screen contains 9 areas or buttons:
Router Signals area
Gate Pick Position area
Gate Place Position area
Centrifuge Contents area
Terminate button
Scale reading area
Gate calibration area
Maintenance area
Log area
Router Signals Area
The system uses the Router Signals checkboxes to display the status of the
sample tube pick and place requests. They are not operator selectable:
Pick Request When checked, indicates a pick request was made.