Centrifuge Module Interface Operating Instructions Introduction
092D0001-02 Rev. A page 29
Using the Sample Manager or the SIQ or manual gate, the operator places
the capped sample tube on the automation system.
• The sample tube arrives at the Centrifuge divert gate, where the bar
code reader identifies the sample tube bar code as a sample to be
centrifuged.
• The divert gate diverts the sample tube to the side track.
• The sample tube moves to the Centrifuge interface gate.
Figure 15: Centrifuge Station Divert and Interface Gates
CAUTION
Do not load glass sample tubes or unsupported sample tube sizes in
Sample Manager trays for processing at the Centrifuge module. Routing a
glass or incorrect sample tube size to the Centrifuge module can result in
breakage, spilled sample, contamination, and system damage. Use only
tubes listed in Appendix C, Specifications.