ARCADIS Orbic Quick Guide
Examination Procedure
33
Collimator setting
For radiation protection reasons collimators
should be inserted. These can be set at the
C-arm system.
(1) Moves the iris diaphragm into/out of the
beam path
(2) Rotates the semitransparent diaphragm
(3) Moves the semitransparent diaphragm
into/out of the beam path
The iris diaphragm is a multileaf collimator
used for concentric collimation of the radia-
tion beam (4).
Activate the iris diaphragm for small
objects, e.g. the wrists (1).
The semitransparent diaphragm acts as a den-
sity compensation. A diaphragm is moved
close to the extremity on both sides. Apart
from reducing radiation, this prevents overex-
posure at the edges of the extremities (5).
Rotate the semitransparent diaphragm
corresponding to the position of the
object (2).
Adapt the semitransparent diaphragm to
the size of the object (3).