68 Image Quality
MAMMOMAT Novation DR SPB7-250.815.06.02.02 Siemens AG
12.05 CS SD 24
Page 68 of 116
Medical Solutions
3. Adjust the grid switch so it will activate and deactivate on either edges of the slits (see
(Fig. 45 / p. 68) and (Fig. 46 / p. 68)). You’ll hear a "click" when the switch activates or
deactivates.
Fig. 45: Adjust grid switch on edge 1
Fig. 46: Adjust grid switch on edge 2
4. Tighten the three screws.
5. Move the grid to the other end position and check that the grid switch activates/deacti-
vates on either edge of the slit (see (Fig. 46 / p. 68) and (Fig. 45 / p. 68)).
Adjusting the "wheel" that is mounted on the grid motor.