70 Image Quality
MAMMOMAT Novation DR SPB7-250.815.06.02.02 Siemens AG
12.05 CS SD 24
Page 70 of 116
Medical Solutions
Fig. 48: Adjusting screws
Adjusting the grid speed
1. Start the service software.
Main menu -> Configuration -> Grid speed
2. Set the parameter as shown in (Fig. 49 / p. 70).
Fig. 49: Adjusting the grid motor speed
3. Check the grid speed with the oscilloscope. The grid speed can be measured with the
oscilloscope. See also Mammomat Novation, System
Start-up...,SPB7-250.815…chapter "X-Ray tube checks", for systems with WH AWS .
Measurement: