Siemens AG SPR8-220.101.04.03.02 MOBILETT Plus HP
07.05 CS PS 24
General 21
Page 21 of 64
Medical Solutions
2.3 Are the used accessories free from
damage?
2.1.1 Damage, check:
SIM DAP (optional)
2.4 Are the system’s radiation protection
devices installed and without any dam-
ages? (no radiation protection of the build-
ing)
2.15 Checking the radiation indicator,
checks:
SIE Radiation indicator
SIE Audible signal
2.16 Checking the manual exposure inter-
rupt, checks:
SIE Message "USE20"
SIE 10 short beep signals
2.5 Are the required operator documents
complete, present and legible?
The operating instructions must be
present. Ask the client for further required
operator documents and check if the docu-
ments are complete, present and legible.
2.6 Are all warning labels in place and rec-
ognizable?
2.1 Visual inspection, check:
SIM Labels
2.7 Are the operating symbols, light indica-
tors and button labels OK?
2.5 Maintenance tests, check:
SIE Display test
2.20 Checking the kV and mAs displays,
checks:
SIE Visibility of displays
SIE Correct displays
3. Electrical checks n.a.
3.1 Measurement of protective conductor
resistance
2.17 Protective ground/ earth measure-
ment, check:
SIE Measured value
3.2 Measurement of leakage current or
equivalent unit leakage current
n.a.
3.2.1 Unit leakage current n.a.
3.2.2 Equivalent unit leakage current 2.18 Equivalent leakage current measure-
ment, check:
SIE Measured value
3.3 Measurement of patient leakage cur-
rent or equivalent patient leakage current
n.a.
3.3.1 Patient leakage current measurement n.a.
Technical Safety Checks,
Test Certificate - check
Maintenance instructions – check/
Comments