SITRANS FM100
page 62 A5E48762204-AA
Index
[hex]
Name
Description
Factory default
Max Value
Min Value
Length
[Bytes]
Data Type
Access
0x0170
OUT2Alarm
Hysteresis
Switching
hysteresis for the
alarmoutput
1,0
9999,0
-9999,0
4
FloatT
R/W
0x0174
OUT2Alarm
SuppressionFactor
How many times
the threshold must
be hit in order to
switch the alarm
output
0
60
0
1
UIntegerT
R/W
0x0175
OUT2Alarm
Suppression
Direction
for which direction
the suppression
factor is used
0
(0) - Up
(1) - Down
(2) - Both
1
UIntegerT
R/W
0x0176
OUT2Analog
NamurStandard
If enabled (1) the
analogoutput
conforms with the
NAMUR Standard
NE42. If disabled
(0) the
analogoutput stays
in his equivalent
range (e.g. 4-
20mA)
1
(0) - NAMUR disabled
(1) - NAMUR enabled
1
UIntegerT
R/W
0x0177
OUT2AnalogValue
0mA
The value from the
slot used for the
0mA scaling point
0,0
OUT2Analog
Value20mA
MRS
4
FloatT
R/W
0x017B
OUT2AnalogValue
4mA
The value from the
slot used for the
4mA scaling point
0,0
OUT2Analog
Value20mA
MRS
4
FloatT
R/W
0x017F
OUT2AnalogValue
20mA
The value from the
slot used for the
20mA scaling point
100,0
MRE
OUT2Analog
Value0mA
4
FloatT
R/W
0x0183
OUT2AnalogValue
0V
The value from the
slot used for the
0V scaling point
0,0
OUT2Analog
Value10V
MRS
4
FloatT
R/W
0x0187
OUT2AnalogValue
2V
The value from the
slot used for the
2V scaling point
0,0
OUT2Analog
Value10V
MRS
4
FloatT
R/W
0x018B
OUT2AnalogValue
10V
The value from the
slot used for the
10V scaling point
100,0
MRE
OUT2Analog
Value0V
4
FloatT
R/W
0x018F
OUT2PulseVolume
The volume
represented by
one pulse
1,0
999,9
0,0
4
FloatT
R/W