SITRANS FM100
page 64 A5E48762204-AA
Index
[hex]
Name
Description
Factory default
Max Value
Min Value
Length
[Bytes]
Data Type
Access
0x01C3
DosingTimeout
Timeout in
seconds for no
flow
0,5
10,0
0,5
4
FloatT
R/W
0x01C7
DosingCounter
Saved dosing
volume counter
stats
0,0
999999,0
-999999,0
4
FloatT
R
Service
0x01CB
ServiceUser
Password
Password for user
service menu and
main menu
0
99999
0
4
UIntegerT
R/W
0x01CF
ServiceUserMenu
Locked
Whether main
menu is locked or
not
0
(0) - not locked
(1) - locked
1
UIntegerT
R/W
0x01D0
SimulationAuto
Stop
Auto stop for
Simulation after
time in minute
10
31
1
1
UIntegerT
R/W
Misc
0x01D2
LanguageSelection
Language selection
0
(0) - English
(1) - German
(2) - French
(3) - Spanish
1
UIntegerT
R/W
0x0262
OperatingHours
Count
Operating hours
counter
0
4294967296
0
4
UIntegerT
R
Flow
0x02F6
CutOff
Cut off for flow
value
0,0
MRE
0,0
4
FloatT
R/W
0x02FA
Unit
Unit used for flow
1
(0) - USER
(1) - L/m
(2) - mL/m
(3) - L/h
(4) - m3/h
(5) - galUS/m
(6) - galUS/h
(7) - galUK/m
(8) - galUK/h
1
UIntegerT
R/W
0x02FB
UserUnit
User Unit used for
flow
1,0
9999,9
0,0
4
FloatT
R/W
0x0314
SimMode
Mode of the
Simulation:
Static, Triangle or
Monotonic
0
(0) - Static
(1) - Triangle
(2) - Monotonic
1
UIntegerT
R/W
0x0315
SimStartValue
Value to start with
the simulation
0,0
9999,0
-9999,0
4
FloatT
R/W

Related product manuals