Loading...
SITRANS FM100
A5E48762204-AA page 65
Index
[hex]
Name
Description
Factory default
Max Value
Min Value
Length
[Bytes]
Data Type
Access
0x0319
SimIncrement
Value
Incrementation
value of the
simulation
10,0
999,0
-999,0
4
FloatT
R/W
0x031D
SimNumber
Intervals
Number of
intervals to
simulation
20
65000
1
2
UIntegerT
R/W
0x031F
SimTimingIntervals
Timinig in ms
between intervals
50
50000
50
2
UIntegerT
R/W
0x0321
ValueInSiUnit
Saved flow value in
SI unit [L/min]
0,0
999999,0
-999999,0
4
FloatT
R
0x0325
MinValueInSiUnit
Saved min flow
value in SI unit
[L/min]
0,0
999999,0
-999999,0
4
FloatT
R
0x0329
MaxValueInSiUnit
Saved max flow
value in SI unit
[L/min]
0,0
999999,0
-999999,0
4
FloatT
R
Volume
0x0359
CountingType
counting type for a
volume slot ->
absolute or
bidirectional
0
(0) - absolute
(1) - bidirectional
1
UIntegerT
R/W
0x035E
Unit
Unit used for
volume
1
(0) - USER
(1) - L
(2) - mL
(3) - m3
(4) - galUS
(5) - galUK
(6) - Barrel
1
UIntegerT
R/W
0x035F
UserUnit
User Unit used for
volume
1,0
9999,9
0,0
4
FloatT
R/W
0x0385
ValueInSiUnit
Saved volume
value in SI unit [L]
0,0
999999,0
-999999,0
4
FloatT
R
Temperature
0x03C2
Unit
Unit used for
temperature
1
(0) - USER
(1) - °C
(2) - °F
1
UIntegerT
R/W
0x03C3
UserUnit
User Unit used for
temperature
1,0
9999,9
0,0
4
FloatT
R/W

Related product manuals