Param‐
eter ID
Meter
version
Display
view
Parameter/data type Factory settings Data range
Fixed parameter or meter data that are not changeable
505 All - Power supply Power supply level Battery or mains power
506 All - Numbers of power-up MLFB-dependent 1 to 4 batteries
507 All - Battery power
508 All - Battery change enable No Yes / No
509 All Se1 Battery installation date PS3 production date and
time
year-month-day T hours:mi‐
nutes:seconds
510 All Se2 Actual battery capacity 100% 100 to 0%
511 All - Communication module type MLFB-dependent No module, RS 485, RS 232, RS
232 always connected
512 All - Number of excitations since
last battery reset
0
513 All - Power status 0 Normal operation,
1: Battery alarm. Actual battery
capacity is below battery alarm
level (% of max capacity)
2: Too low power (enters stand
by mode)
3: As value 1 and 2 together
4: External power gone
5: As value 1 and 4 together
6: As value 2 and 4 together
7: As value 1 and 2 and 4 to‐
gether
514 All - Transmitter temperature Actual degree celsius
540 All - Electrode impedance A Measured values 0 to 185 000 ohm
541 All - Electrode impedance B Measured values 0 to 185 000 ohm
542 All - Low medium impedance
alarm
0 0 to 2.15x10
9
550 All - Coil current disable No Yes / No
551 All - Fixed flow mode enable No Yes / No
Value is reset to "No" in 4 hours.
552 All - Fixed flow value 0 -1x10x10
9
to 1x10x10
9
553 All - High flow alarm limit 1 000 000 000 0 to 1x10
9
554 All - Reverse flow alarm limit -1x10
9
-1x10
9
to 1x10
9
560 All - Repair checksum No Yes / No
Value is reset to "No" when up‐
loading device parameters to pc
after command execution
570 All - Device Product ID vendor id = 42
product id = 27
Siemens MAG 8000 product ID:
vendor id = 42
product id = 27
Appendix
C.2 Parameter lists
SITRANS MAG 8000
146 Operating Instructions, 11/2018, A5E03071515-AE