Param‐
eter ID
Meter
version
Display
view
Parameter/data type Factory settings Data range
Fixed parameter or meter data that are not changeable
590 All with
FW3.07
and lat‐
er
- Raw velocity Measured value
592 All with
FW3.07
and lat‐
er
- Electrode potential Measured value within ±1.2V
C.2.7 600-799
Param‐
eter ID
Meter
version
Display
view
Parameter/data type Factory settings Data range
Fixed parameter or meter data that are not changeable
600 All - Log interval Monthly Daily, Weekly (7 days), Monthly
601 All - Delay of weekly logging from
current day to first day logging
at 00:00:00
0 0 to 30
602 All - - Limit for too high consump‐
tion
1 000 000 -1x10
9
to 1x10
9
603 All - Limit for too low consumption 0 -1x10
9
to 1x10
9
610* All L1 Date of latest log period 2000-01-01 T 00:00:00 year-month-day T hours:mi‐
nutes:seconds
611* All L1 Latest Log period totalized (1)
612* All - Latest Log period totalized (2) 0
613* All - Latest Log period fault status* No faults Active faults in log period;
1: Insulation error
2: Coil current error
3: Amplifier overload
4: Database checksum error
5: Low power warning
6: Flow overflow warning
7: Pulse A overload warning
8: Pulse B overload warning
9: Consumption interval warning
10/L: Leakage warning
11/E: Empty pipe warning
12/C: Low impedance/ high con‐
ductivity warning
13/d: High flow limit warning
14/15/16: Not used
If none of the errors have been
logged, the message is "No
faults"
Appendix
C.2 Parameter lists
SITRANS MAG 8000
Operating Instructions, 11/2018, A5E03071515-AE 147