Example:
A flow velocity of 3 m/s (10 ft./sec.) (V) in a sensor with a diameter reduction from DN 100 to
DN 80 (4" to 3") (d
1
/d
2
= 0.8) gives a pressure drop of 2.9 mbar (0.04 psi).
4.1.3 Mounting the sensor
1. Install gaskets.
2. Ensure connection flange has a smooth surface and is in line with sensor.
Gaskets are recommended but not included in water meter delivery.
Advice for gasket selection:
Only use flat rubber gaskets.
Thickness 1 to 6 mm (0.0 to 0.02 ft) dependent on gap/tolerance.
Inner diameter must be larger than bore of water meter.
Material should be compatible with process fluid.
Hardness should be maximum 75 Shore A.
Maximum allowable torques
Standard bolts must be well lubricated and tightened evenly around gasket.
Leakage/damage to water meter or piping may arise if bolts are overtightened.
Torque calculations
Installing/Mounting
4.1 Sensor installation
SITRANS MAG 8000
22 Operating Instructions, 11/2018, A5E03071515-AE