Guidelines for selection of sensor
Normally the sensor is selected so that V lies within the measuring range 1 to 2 m/s (3 to 7 ft/
sec.).
Min. measuring range: 0 to 0.25 m/s (0 to 0.8 ft/sec.)
Max. measuring range: 0 to 10 m/s (0 to 33 ft/sec.)
Flow velocity calculation formula
Metric measures:
V = 1273.24 x Q / Di
2 (
V: [m/s]; Q: [L/s]; Di: [mm])
or
V = 353.68 x Q / Di
2 (
V: [m/s]; Q: [m
3
/h]; Di: [mm])
Imperial measures:
V = Q x 0.408 / (Pipe I.D.)
2 (
V: [ft/s]; Q: [GPM]; Pipe I.D.: [inch])
or
V = Q x 283.67 / (Pipe I.D.)
2 (
V: [ft/s]; Q: [MGD]; Pipe I.D.: [inch])
C.4 Certificates
You can find certificates on the Internet at AUTOHOTSPOT or on an included DVD.
See also
Certificates on the internet (
http://support.automation.siemens.com/WW/view/en/
10806951/134200)
C.5 Spare parts/Accessories
C.5.1 Ordering of spare parts