C.2.3 200-299
Param‐
eter ID
Meter
version
Display
view
Parameter/data type Factory settings Data range
Fixed parameter or meter data that are not changeable
200 All O4 Fault status No faults 0 to 8191, binary presented with
information 1 for bit 0
1: Insulation error
2: Coil current error
3: Amplifier overload
4: Database checksum error
5: Low power warning
6: Flow overflow warning
7: Pulse A overload warning
8: Pulse B overload warning
9: Consumption interval warning
10/L: Leakage warning
11/E: Empty pipe warning
12/C: Low impedance (high con‐
ductivity) warning
13/d: High flow limit warning
14/A: Reverse flow warning
If no error/warning is pending,
the message is "No faults"
201 All - Alarm configuration list 254 = Alarm 2 to 8 enabled 0 to 8191, See 200
If no alarm is enabled, the mes‐
sage is "None enabled"
202 All - Date of fault log reset PS3 production date and
time
year-month-day T hours:mi‐
nutes:seconds
203 All O4 Non optimal measure time, in‐
cluding the time of device run‐
ning with insulation test fault,
coil current fault, amplifier
fault, database checksum
fault, flow overload fault, or
empty pipe fault.
0
204 All - Reset the fault log and faults No Yes / No
Value is reset to "No" when up‐
loading device parameters to pc
after command execution
205 All - Call up acknowledge No Yes / No
Value is reset to "No" when up‐
loading device parameters to pc
after command execution
206 All - Battery alarm level 10% 0 to 100%