Power supply
12-24 V AC/DC power supply Input voltage range 12/24 V AC/DC (10 ... 32 V DC)
Power consumption 2 VA
Isolation Class II
Fuse 1000 mA T - Not replaceable
Short circuit protection Module is protected from short circuit on the output con‐
nector. Both during mains and backup supply
Conformity IEC 61010-1
OIML R 49-1
EMC:
EN 61000-6-3
EN 61000-6-2
115 to 230 V AC mains supply Input voltage range 115 to 230 V AC, +15% to −20%, 50-60 Hz
Power consumption 2 VA
Isolation Class II
Fuse 250 mA T - Not replaceable
Short circuit protection Module is protected from short circuit on the output con‐
nector. Both during mains and backup supply
Conformity IEC 61010-1
OIML R49-1
EMC:
EN 61000-6-3
EN 61000-6-2
Input cable for
12/24 V AC/DC and
115 to 230 V AC power supply
Factory-mounted PUR cable 2 x 1 mm
2
(brown, blue)
Length = 3 m
Resistance Sunlight and water
Outer diameter 7 mm (0.28")
Rated voltage 300 ... 500 V AC
Testing voltage 2000 V AC
Temperature range Fixed laying:
Flexible application:
−40 to +90 °C (-40 to 194 °F)
−30 to +80 °C (-22 to 176 °F)
Bending radius Min. 28 mm (fixed installation)
Pulling force Max. 200 N
Output Female connector
Backup battery Male connector
1)
Lithium batteries are subject to special transportation regulations according to United Nations
"Regulation of Dangerous Goods, UN 3090 and UN 3091". Special transport documentation
is required to observe these regulations. This may influence both transport time and costs.
10.5 Modbus RTU
Siemens Flow Instruments Modbus RTU specification for add-on modules
Device type Slave
Baud rates 1200, 2400, 4800, 9600, 19 200, 38 400 bits/sec.
Technical data
10.5 Modbus RTU
SITRANS MAG 8000
Operating Instructions, 11/2018, A5E03071515-AE 95