6 Operating Instructions
30 Minilab man-CH-PVF-C-V17-2
6.1.2.4 Scale Range
The measuring range is set with this menu. Because a change in the
measuring range also requires an alteration of the measuring amplifier,
this entry is protected by password and only accessible by the technical
service of Simpson Technologies.
6.1.2.5 Force Test Sample
The Force Test sample is set to 1963N by default. This corresponds to a
pressure of 100 N/cm
2
on the test specimen.
The Force Test sample can be changed with customer username C”
and password 3472.
The change of the Force Test sample has a direct influence on
the subsequent strength measurement: the higher the force, the
higher the compressive strength. Before changing the force, the
influence of the compactability should also be examined (with the
same Force Test sample).