44
401925/C
Verifyingthecommunicationwitha
synchronizationsystem
Whenevermorethanonehydroacousticsystemisinstalledonavessel,interference
mayoccur.TheEK80offersfunctionalityforremotetransmitsynchronization.Itcan
besetuptooperateineitherMasterorSlavemode.Synchronizationisrequiredin
ordertoavoidinterferenceiftheEK80isusedsimultaneouslywithotherhydroacoustic
instrumentswithinthesamefrequencyrange.YoudonotneedtodothistestiftheEK80
shallonlyoperateinStandalonemode.
Prerequisites
TheEK80isinstalledasspeciedintheEK80Installationmanual.Tomakesurethat
thetheinterfaceisfunctional,arelevantsynchronizationsystemmustbeconnected
totheEK80.
TheEK80systemisturnedonandoperatesnormally.
For"slave"operation,aremotesystem(forexampleK-SyncorSimradTU40)must
beavailabletoprovidetriggerpulses.
For"master"operation,aremotehydroacousticsystem(sonar,echosounder)is
connected.Thisremotesystemmustbesetupin"slave"mode.
Thevesselisberthed.
Neithertoolsnorinstrumentsarerequired.
Context
Whenevermorethanonehydroacousticsystemisinstalledonavessel,interferencemay
occur.Toavoidinterference,youhavetheseoptions:
Thesystemsareallconnectedtoacommonsynchronizationsystem.
Oneoftheacousticsystemsissetupas"master",andcontrolsthetransmissionson
theothersystems.
Procedure
1TestthesynchronizationwhenEK80operatesinMastermode.
aOntheSetupmenu,selectInstallation.
bSelectSynchronization.
cRecordthecommunicationparametersintheresulttable.
dSelectsynchronizationmode.
eSelectthesynchronizationdelay.
fMakesurethatrelevanthydroacousticsystemsconnectedtotheEK80are
synchronized.
2TestthesynchronizationwhenEK80operatesinSlavemode.