5 Putting the system into operation Sirona Dental Systems GmbH
Operating Instructions C4
+
59 06 313 D 3382
22 D 3382.201.01.13.02
Display of next maintenance date
the next maintenance date is less than 30 days away,
this message appears each time the unit is switched on.
(see MAINTENANCE Mode dialog).
After this date has expired, the following message is dis-
played:
MAINTENANCE
REQUIRED
The display message then disappears as soon as an
instrument is removed from its holder.
NOTICE
i
This message can be reset only by the service techni-
cian after completion of maintenance work.
CAUTION
The dental chair has a maximum load capacity of 135kg
according to EN ISO 6875 (tested with a four-fold safety
factor according to IEC 60601-1).
The patient’s arms and legs must rest on the upholstery
of the chair!
Preselecting the user
The treatment center offers the possibility of managing
two different chair and instrument programs for two
users.
Preselect user A or B using the A/B key, with all instru-
ments in place.
The preselected user is shown in the top left corner of
the display. All settings which have been stored for that
user are then activated.
MAINTENANCE
IN 30 DAYS
wrong
right
A/B
READY
10:04:27
A
B