13 Water unit Sirona Dental Systems GmbH
13.4 Disinfection system (optional) Operating Instructions C4
+
59 06 313 D 3382
70 D 3382.201.01.13.02
13.4 Disinfection system (optional)
When the message DESINF flashes on the dentist ele-
ment display during treatment, you should refill 1 liter of
DENTOSEPT P at the latest after the end of the treat-
ment day.
When the message flashes, it is not possible to perform
sanitation.
Remove the cover behind the cuspidor and fill in 1 liter
of DENTOSEPT P.
The filler trough should always be kept clean. Always
make sure that no tap water enters the filler trough!
For reorders:
DENTOSEPT P:
1 carton = 6 x 1 liter, REF 33 18 156
For Canada and Japan only:
1 carton = 6 x 1 liter, REF 59 69 535
Microbiological water test (after an operating pause of
more than 1 week)
Sanitation of the unit
Perform sanitation every 4 weeks. This is the only way
to effectively combat the formation of biofilm in the water
paths (see instructions “Care and Cleaning by the
Practice Team”).
DESINF
A
RPM