11 Basic treatment center settings in the Mode dialog Sirona Dental Systems GmbH
11.19 Mode: WHITESCREEN YES – NO Operating Instructions C4
+
59 06 313 D 3382
64 D 3382.201.01.13.02
11.19 Mode: WHITESCREEN YES – NO
Presetting a white screen for the SIVISION
monitor
If YES flashes –
The “X-ray image viewer” key switches ON/OFF both
the X-ray image viewer and a WHITE SCREEN on the
SIVISION 3 monitor.
If NO flashes –
The “X-ray image viewer” key only switches the X-ray
image viewer ON and OFF.
If an X-ray image viewer and SIVISION 3 are installed,
always select “NO”.
11.20 Mode: A/B VIDEO MODE ON – OFF
Locking the video mode switchover
This mode is only displayed in connection with an inte-
grated SIVISION 3.
If ON is flashing –
You can use the “A/B” key to switch between VIDEO
mode and PC mode after picking up the SIVISION 3
camera.
If OFF is flashing –
You cannot use the “A/B” key to switch between VIDEO
mode and PC mode for SIVISION 3.
This prevents the selected mode from being changed
inadvertently.
11.21 Mode: BLOW OUT SPRAY
Prevents drop formation
You can activate the instrument blow-out function in this
setting option. This removes any residual water (one to
two drops) by blowing air through the instrument heads.
Just press the
– or + key to switch this function ON or
OFF.
The selected item flashes.
NOTICE
i
This function is switched OFF in the factory setting.
NOTICE
i
Sirona recommends activating this function especially
for the C3
+
treatment center.
WHITESCREEN
YES – NO
A / B VIDEO MODE
ON – OFF
A/B
BLOW OUT SPRAY
YES – NO