13.3 Amalgam rotor
59 06 313 D 3382
D 3382.201.01.13.02
69
13.3 Amalgam rotor
The message AMALG appears on the display when the
amalgam rotor needs replacement, i.e., when having
reached a filling level of 95% (see instructions “Care
and Cleaning by the Practice Team”).
A continuous sound indicates that the rotor is full and
no longer functioning properly.
An intermittent sound indicates a system error. The
water supply is interrupted. Please contact your service
engineer.
NOTICE
i
The rotor inside the amalgam separator must be re-
placed at least once a year, even if the AMALG message
does not light up. Otherwise, unpleasant odors can de-
velop.