No Air Supply Air supply is too low (<2.0 bar) to
assure normal operation.
Non-latching HIGH priority alarm.
Suspends high oxygen alarm.
Priority: HIGH
Type: TECHNICAL
Device will deliver 100%
oxygen as set flow.
Connect air supply. If
ventilating - remove patient
to an alternative form of
ventilation. Remove the unit
from service.
No Oxygen Supply Oxygen supply is too low (<2.0 bar)
to assure normal operation. Only
active if FiO
2
is set >21%.
Non-latching HIGH priority alarm.
Suspends low oxygen alarm.
Priority: HIGH
Type: TECHNICAL
Device continues to deliver
air only gas to patient.
User can prevent
recurrence by setting FiO
2
to 21%.
Connect oxygen supply.
If no oxygen present, set
the unit to run on air only by
canceling the alarm
message.
If ventilating - remove
patient to an alternative
form of ventilation. Remove
the unit from service.
High BPM rate The measured breath rate exceeds
90 BPM for at least 6 seconds
when the Apnoea is set in the
range of 5 and 90 seconds.
Disabled if Apnoea is OFF (--)
High rate could indicate false
breath detection due to pressure
signal noise (e.g. from water in
breathing circuit). False detection
risk masking an apnoea condition.
Non-latching HIGH priority alarm.
Priority: HIGH
Check Patient, check
patient circuit. Adjust
Apnoea parameters.