Zero Pressure Measured CPAP is < 0.6 mbar,
while a flow is being delivered and
safety cut out is not in progress.
Indicates a sensor fault or sensor
line disconnect.
Non-latching HIGH priority alarm.
Priority: HIGH
Type: TECHNICAL
Fresh gas is set to fixed flow
of 6 l/min, until CPAP >0.9
mbar is restored.
Check Patient, check
patient circuit.
High Leak.
(Flow mode)
High leak alarm in triggered when
delivered pressures goes below
the captured pressure by more
than 2mbar and remains active for
10 seconds.
Note: If the measured value is 2
mbar, the alarm operates for
pressures less than 1 mbar.
Possibly the circuit or patient
interface has become
disconnected.
Non-latching HIGH priority alarm.
Cannot be disabled.
Priority: HIGH
Type: TECHNICAL
Set alarm threshold. Check
patient, check patient
circuit.
Alarm message Description Action