Alarm message:
Gas Flow Error
Unit started with complete circuit
attached.
Disconnect proximal airway
line and cycle power.
Alarm message:
High BPM rate
Water in circuit creating circuit noise.
See “High BPM rate” on page 100.
Check patient circuit.
Adjust trigger sensitivity.
Symptom Possible Cause Remedy