HRX-OM-W002-A
Chapter 5 Display and Setting of Various Functions
HRR Series 5.5 Advanced setting mode
Ambient temperature lower limit alarm
17. Press the [▼] key. Displays screen for setting the lower limit of the ambient
temperature alarm.
Displayed for air-cooled type only. (This item is not displayed for water-cooled type)
Lower limit of alarm AL11: Ambient temperature alarm can be changed.
●Setting range: 5.0 to 40.0℃ (initial value: 10.0℃)
Switching of the leakage alarm
18. Press the [▼] key. Displays screen for setting the leakage alarm operation.
Operation for alarm AL04: Leakage alarm can be selected.
Operation of this product will continue when the
alarm occurs.
Operation of this product will stop when the alarm
Occurs.
Switching of the contact input signal 1
19. Press the [▼] key. Displays screen for setting the operation of the Contact input signal
1.
Operation for alarm AL25: Contact input signal 1 detection can be selected.
Operation of this product will continue when the alarm
occurs.
Operation of this product will stop when the alarm occurs.
AMB ⇒ LOTMP (Alternately displayed)
Operation of leakage alarm
Operation of the contact input signal 1
INP1 ⇒ ALARM(Alternately displayed)
WATER ⇒ LERK ⇒ ALARM(Alternately displayed)
Ambient temperature lower limit alarm