EasyManua.ls Logo

SMC Networks PFCA7 Series - Reference Condition Setup; Display Unit Configuration

SMC Networks PFCA7 Series
83 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
-25-
No.PF※※-OMA1002
(2) Reference condition settings
Press the SET button.
STD: Standard condition (Flow rate converted into volume at a temperature of 20° [C],
absolute pressure of 101.3 [kPa], and humidity of 65 [%R.H.])
NOR: Normal condition (Flow rate converted into volume at a temperature of [C],
absolute pressure of 101.3 [kPa], and humidity of 0 [%R.H.])
Press the SET button.
Returns to the [F0 System] settings.
(3) Display unit settings
Press the SET button.
: A product with no units selection function (fixed to SI units) does not indicate the display unit [cfm].
Press the SET button.
Returns to the [F0 System] settings.
Press the UP or DOWN button
to select an item.
Select [Fluid cond].
Press the UP or DOWN button
to select an item.
Select [Unit].
2023-12-20 14:49
DL145266

Related product manuals