OPERATING MANUAL PAT-1, PAT-2, PAT-2E, PAT-10 version 1.02
nominal current: min 1 mA ... 1.4 mA
adjustable lower limit within the range of 0,1 MΩ...99,9 MΩ with resolution of 0,1MΩ
adjustable measuring time: 3s…180 s with a resolution of 1s
detection of a dangerous voltage before commencing a measurement
discharging the object tested
Note: For R <0,25 MΩ, the uncertainty is not specified.
Measurement of insulation resistance using test voltage of 500V
Test range according to IEC 61557-2 for UN = 500V: 0,50 MΩ…99,9 MΩ
Measurement uncertainty
(basic)
0 % for R ≤ 20 MΩ
Unspecified for R > 20 MΩ
Accuracy of generated voltage (Robc [] 1000*U
N
[V]): -0 % +30 % from the set value
nominal current: min 1 mA ... 1.4 mA
adjustable lower limit within the range of 0,1...99,9 MΩ with resolution of 0,1MΩ
adjustable measuring time: 3s…180s with a resolution of 1s
detection of a dangerous voltage before commencing a measurement
discharging the object tested
Note: For R < 0,50 MΩ, the uncertainty is not specified.
Measurement of substitute leakage current
Measurement uncertainty
(basic)
opening voltage: 25 ... 50V
internal resistance of the testing device 2kΩ ± 20%
adjustable upper limit in the range of: 0,01 mA…19,9 mA with resolution of 0.01mA/0.1mA
adjustable measuring time in the range of: Cont , 5…180s with resolution of 1 s