EasyManua.ls Logo

Sony RCD-W10 - Page 40

Sony RCD-W10
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40
RCD-W10
5) As prompted by the message shown in Fig. 4-24, check that
the waveform as shown in Fig. 4-23 appears on an oscilloscope.
6) As prompted by the message shown in Fig. 4-26, check that
the waveform as shown in Fig. 4-25 appears on an oscilloscope.
Note :The clear RF signal waveform means that the shape “ ” can be clearly distinguished at the center of the waveform.
RF signal waveform
VOLT/DIV : suitable value
TIME/DIV : 500ns
4 Fig. 4-22
RF signal waveform
VOLT/DIV : suitable value
TIME/DIV : 500ns
4 Fig. 4-23
RF signal waveform
VOLT/DIV : suitable value
TIME/DIV : 500ns
4 Fig. 4-25
4 Fig. 4-24
4 Fig. 4-26
7) Check that the message “CDR PB-CAL completed!” appears.
NG Judgment Result Indication
4-2-1. Playback Measurement NG!
Contents : This message appears when the BLER (Block Error Rate) during playback is outside the specification.
Measure : Check if there is any abnormality in RBD1.
Replace KRM-220CAA.

Other manuals for Sony RCD-W10

Related product manuals