EasyManua.ls Logo

Soredex CRANEX Novus e - L1800 Generator Board

Soredex CRANEX Novus e
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
46 User manual 208898
6. Troubleshooting and Maintenance CRANEX Novus e
E16
REASON
Z-motor has overheated.
SOLUTION
Wait until the Z-motor, driving the unit up and down, has cooled down. The error
message clears automatically when the motor has cooled down to the appropriate
level. If the error message keeps on appearing or does not disappear within a
reasonable amount of time, contact service.
E19
REASON
Exposure switch stuck down during unit start.
SOLUTION
Switch the unit off and check that the exposure switch is not stuck in the exposure
position. Switch the unit on again. If the message reappears, contact service.
Exx (all other E errors except E4, E5, E16 and E19.
SOLUTION
Clear the error message and try to take an exposure without a patient. If the error
message reappears, switch the unit off, wait for half a minute and then switch the
unit on again. If the error message reappears contact service.
Approved: Ukkonen Juha-Pekka 2015-03-13 13:08
Reviewed: Savinen Antti Juhani 2015-03-12 11:32
Approved
See PDM system to determine the status of this document. Printed out: 2016-01-27 12:19:26
D508980, 4
Copyright © 2015 by PaloDEx Group Oy. All rights reserved.

Table of Contents

Other manuals for Soredex CRANEX Novus e

Related product manuals