CRANEX Novus e 6. Troubleshooting and Maintenance
User errors
H1
REASON
The exposure button was released during an exposure.
SOLUTION
Clear the error message and check if the attempted exposure is suffi cient for the
diagnostic task. If it is not, take a new exposure.
If the exposure failed while the exposure button was still being pressed, check the
exposure switch by taking a test exposure without patient to see if the exposure
button is defective or not. If the same problem occurs again, contact service.
System errors
E4
REASON
Tubehead too hot or too cold.
SOLUTION
When the error message automatically clears the tubehead has reached the
correct operating temperature. In normal conditions this will take about 30 minutes
for the tubehead to reach the correct temperature. If the error message does not
disappear within a reasonable amount of time, contact service.
E5
REASON
Line voltage not within limits.
SOLUTION
If the error message reappears it indicates that the voltage is not within limits.
The error message will automatically clear when the voltage returns to the correct
level. If the error message keeps on appearing or does not disappear within a
reasonable amount of time, contact service.