User manual 208898 49
CRANEX Novus e 6. Troubleshooting and Maintenance
Yearly maintenance
Once a year an authorized service technician must
carry out a full inspection of the unit. The following
tests and checks must be carried out:
– a kV/mA test
– a beam alignment test
– check that the safety ground is connected
– check that the positioning lights operate
– check that no oil is leaking from the tube head
– check that the power lead is not damaged in any
way.
– check that all covers and mechanical parts are
correctly secured and have not come loose.
– check that any vents in the covers are not
blocked with dust and that no dust has
accumulated inside the unit.
A full description of all the tests and checks is
described in the Service Manual.
Approved: Ukkonen Juha-Pekka 2015-03-13 13:08
Reviewed: Savinen Antti Juhani 2015-03-12 11:32
Approved
See PDM system to determine the status of this document. Printed out: 2016-01-27 12:19:26
D508980, 4
Copyright © 2015 by PaloDEx Group Oy. All rights reserved.