User manual 208898 49
CRANEX Novus e 6. Troubleshooting and Maintenance
Yearly maintenance
Once a year an authorized service technician must
carry out a full inspection of the unit. The following
tests and checks must be carried out:
a kV/mA test
a beam alignment test
check that the safety ground is connected
check that the positioning lights operate
check that no oil is leaking from the tube head
check that the power lead is not damaged in any
way.
check that all covers and mechanical parts are
correctly secured and have not come loose.
check that any vents in the covers are not
blocked with dust and that no dust has
accumulated inside the unit.
A full description of all the tests and checks is
described in the Service Manual.
Approved: Ukkonen Juha-Pekka 2015-03-13 13:08
Reviewed: Savinen Antti Juhani 2015-03-12 11:32
Approved
See PDM system to determine the status of this document. Printed out: 2016-01-27 12:19:26
D508980, 4
Copyright © 2015 by PaloDEx Group Oy. All rights reserved.