EasyManua.ls Logo

Soredex CRANEX Novus e - CCD Sensor; Filter Board

Soredex CRANEX Novus e
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A-1
CRANEX Novus e Appendix A. Technical Information
Appendix A. Technical Information
A.1 Technical speci cations
Type
CRA-2
Classi cation
Complies with
- IEC 60601-1
- IEC 60601-1-2
- IEC 60601-1-3
- IEC 60601-1-4
- IEC 60601-2-7
- IEC 60601-2-28
- IEC 60601-2-32
- UL 60601-1
Conforms with the regulations of DHHS Radiation Performance Standard, 21CFR
Subchapter J.
Safety according to IEC 60601-1
Protection against electric shock - Class 1
Degree of protection - Type B applied with no conductive connection to the patient
Protection against the ingress of liquids - IP 20
Disinfection methods:
- Mild soapy water (non-abrasive)
- Non-alcohol based disinfectant for the chin rest
- Disposable plastic covers for bite piece, chin rest and lip support
For use in environments where no ammable anaesthetics nor ammable cleaning
agents are present
Mode of operation - continuous operation/intermittent loading
Unit description
A dental panoramic x-ray unit with a high frequency switching mode x-ray
generator. The unit takes panoramic exposures.
The unit uses a CCD sensor as an image receptor.
Generator
TUBE
Toshiba D-052 SB or D-054 SB or CEI 105/5
TUBEHEAD HOUSING ASSEMBLY
THA-M-X (Where X is the version)
FOCAL SPOT
0.5 mm (IEC 60336/2005)
Approved: Ukkonen Juha-Pekka 2015-03-13 13:08
Reviewed: Savinen Antti Juhani 2015-03-12 11:32
Approved
See PDM system to determine the status of this document. Printed out: 2016-01-27 12:19:26
D508980, 4
Copyright © 2015 by PaloDEx Group Oy. All rights reserved.

Table of Contents