A-2
Appendix A. Technical Information CRANEX Novus e
FOCAL SPOT ACCURACY
The accuracy is 10 mm from the marking on the tubehead cover
TARGET ANGLE
TARGET MATERIAL
Tungsten
NOMINAL X-RAY TUBE VOLTAGE WITH THE HIGHEST X-RAY TUBE CURRENT
100 V~ 77 kV at 8 mA
120 V~ 77 kV at 10 mA
220-240 V~ 77 kV at 10 mA
X-RAY TUBE VOLTAGE tolerance is ± 4 kV
HIGHEST X-RAY TUBE CURRENT WITH THE HIGHEST X-RAY TUBE VOLTAGE
100 V~ 8 mA at 77 kV
120 V~ 10 mA at 77 kV
220-240 V~ 10 mA at 77 kV
X-RAY TUBE CURRENT tolerance is ± 1 mA
NOMINAL ANODE INPUT POWER AND NOMINAL ELECTRIC POWER
100 V~ 616 W at 77 kV, 8 mA, 9 s
120 V~ 770 W nominal at 77 kV, 10 mA, 9 s
220-240 V~ 770 W nominal at 77 kV, 10 mA, 9 s
X-RAY TUBE VOLTAGE STEPS
100 V~ 63, 66, 70, 73 and 77 kV
120 V~ 63, 66, 70, 73 and 77 kV
220-240 V~ 63, 66, 70, 73 and 77 kV
X-RAY TUBE VOLTAGE tolerance is ± 4 kV
X-RAY TUBE CURRENT
100 V~ 8 mA
120 V~ 10 mA
220-240 V~ 10 mA
X-RAY TUBE CURRENT tolerance is ± 1 mA
Approved: Ukkonen Juha-Pekka 2015-03-13 13:08
Reviewed: Savinen Antti Juhani 2015-03-12 11:32
Approved
See PDM system to determine the status of this document. Printed out: 2016-01-27 12:19:26
D508980, 4
Copyright © 2015 by PaloDEx Group Oy. All rights reserved.