24 User manual 208898
3. Taking an Exposure CRANEX Novus e
Temporomandibular Joint (TMJ)
1. Slide the lip support on to the support holder.
Use the slot nearest the mirror for patients with
large skulls and the slot furthest away from the
mirror for patient with small skulls.
2. Place the disposable cover on to the lip support.
3. Press the TMJ key to select TMJ program. The
key color will change which indicates that the
program has been selected.
The magnifi cation is 1.25.
4. A default kV value for the selected program will
appear on the display.
If you consider that the default kV value is not
correct for the patient being examined select a
different kV value based on the patient’s size,
age and estimated bone density.
To change the kV value press the kV plus (+) or
minus (-) key.
The recommended values are:
- Large Adult - 77 kV
- Adult - 73 kV
- Juvenile or small adult - 70 kV
- Child - 63 or 66 kV
Approved: Ukkonen Juha-Pekka 2015-03-13 13:08
Reviewed: Savinen Antti Juhani 2015-03-12 11:32
Approved
See PDM system to determine the status of this document. Printed out: 2016-01-27 12:19:26
D508980, 4