7. Quality Assurance CRANEX Novus e
50 User manual 208898
7. Quality Assurance
Additional Quality Assurance (QA) exposures
can be taken after installation and periodically to
ensure that the image quality remains stabile. The
QA exposures can be taken using a special QA
program.
To take QA images you need an additional Cu
(Copper) fi lter, Digital Test Tool (Prüfkörper test tool)
and Ball and Pin phantom that need to be ordered
separately.
There are different thicknesses of the Cu-fi lters,
select the correct one depending on the national
requirements.
The unit provides two different QA program
selections (QA1 and QA2):
- QA1 is intended for use with 0,8mm thick Cu fi lter
(in accordance to IEC61223-3-4)
- QA2 is intended for use with 1.8mm thick Cu fi lter.
Approved: Ukkonen Juha-Pekka 2015-03-13 13:08
Reviewed: Savinen Antti Juhani 2015-03-12 11:32
Approved
See PDM system to determine the status of this document. Printed out: 2016-01-27 12:19:26
D508980, 4