User’s manual 208898 57
Cranex Novus e 8. Warnings and precautions
enough to fully protect the user from radiation.
The radiation protection area must be at least 60
cm wide, 60 cm deep and 200 cm high.
The radiation protection screen must include
lead shielding with a minimum thickness 0.5
mm and the screen must conform to all national
regulations concerning radiation shielding of
dental/medical devices.
• The user must be able to see and hear the
patient during an exposure.
• The user must be positioned so that he/she
can see the exposure warning light/indicator
and hear the audio exposure warning signal
during the exposure. If the unit is installed in
such a place where the exposure warning light/
indicator cannot be seen or the audio exposure
warning signal cannot be heard, then a separate
exposure warning light/indicator device should
be used. Please contact your local service for
help.