EasyManua.ls Logo

Söring Arco-3000 - Block Diagram Cut Module

Söring Arco-3000
89 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
25/08/2004 ©
Söring
Service-Manual Arco- and RF-units
1
not valid for A-3000/ -2000
2
not valid for Arco-2000
-37-
3
not valid for MBC600
4
not valid for Arco-1000
Block diagram Cut module
range
selection
driver
Multiplier
UxI
Oscillator
ca. 350kHz
output stage
(MOS-bridge)
filter
power
control
range
selection
current
control
current/voltage-
detection
Pulse-
oscillator
Pulse width
overtone-
detection
electric arc-
border value
electric arc-
control
Logic
cycle control
Interface
Businterface
analog / digital
ON(1)
PWR_SEL(3)
CURR_SEL(1)
MONO/BIPO(1)
CO/CUT(1)
PULSE(1)
PW_SET(A)
I_SET(A)
FRQ_SET(A)
DUTY_SET(A)
DOWN
PW_REG
SP_REG
ON
ON
BR_ENABLE
Mono/Bipo
Mono/Bipo
output transformer
HF_OUT
U
I
PWR_SEL
PW_SET HOLD
ON
PW_REG
Interface
Mono/Bipo
FRQ_SET
DUTY_SET
HOLD
BR_ENABLE
SP_SET
U
I
CURR_SEL
I_SET
ON
HOLD
DOWN
DOWN
to Interface
SP_REG
Interface

Table of Contents

Related product manuals