25/08/2004 ©
Söring
Service-Manual Arco- and RF-units
1
not valid for A-3000/ -2000
2
not valid for Arco-2000
-68-
3
not valid for MBC600
4
not valid for Arco-1000
Adjustment resonance frequency monopolar:
- Start equipment in service mode , select service function 11, connect RF-meter to
monopolar handpiece with 100
load, press handpiece switch "Cut" and adjust by
means of P3 on cut module to maximum output power. Power approx. 85W.
Adjustment resonance frequency bipolar:
- connect bipolar connecting cable to RF-meter, select 100
load, start the
equipment (foot switch cut). Adjust by means of P4 to maximum output power.
Power approx. 85W.
Adjustment offset of power measuring amplifier
- Select service function 12, sub-function 1 (Power supply delivers no voltage,
selected power "zero", adjust potentiometer P2 so that LED D47 (cut module) just
starts to light up. Press "Cut"-switch with this adjustment.
Check of arc regulation
- Select service function 12 (Power supply delivers no power, select appropriate
parameters), connect signal generator (sine signal source resistance 50
) to
handpiece and PE-electrode. Set frequency to 1.05 MHz. Increase output voltage
until LED D31 (cut module ) just starts to light up. The required output voltage must
be in the range 2 to 5V
ss
. Press "Cut"-switch with this adjustment.
12.5.4 Adjustment of RF-Out B2 (SMD) [from SW 3.10]
Necessary equipment
- RF-Handpiece with two fingerswitches
- Patient electrode
Adjustment of Offset-Voltage
- start unit in Service-Mode
- choose Service function 15, sub function 3
- Monopolar Handpiece and Patient electrode have already to be plugged in
- Unit has to be switched on for at least 10 minutes already!
- start unit periodically with handpiece button CUT (yellow) or Coag. (blue) (Only
this way the offset-value will be displayed in the Coag.-field correctly).
- adjust with poentiometer P1 on HFO-B2 (piggyback) a value of 25-30
- Leave Service-Mode and start unit again