EasyManua.ls Logo

Söring Arco-3000 - Power Diagram Force-Coagulation -Effect 9; Power Diagram Soft-Coagulation

Söring Arco-3000
89 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
25/08/2004 ©
Söring
Service-Manual Arco- and RF-units
1
not valid for A-3000/ -2000
2
not valid for Arco-2000
-80-
3
not valid for MBC600
4
not valid for Arco-1000
Power diagram Force-Coagulation -Effect 9
4
0
20
40
60
80
100
120
140
100 150 200 400 600 1000 2000
Resistance in Ohm
Power in Watt
Force 1 (Eff 1)
Force 2 (Eff 1)
Force 3 (Eff 1)
Force 4 (Eff 1)
Force 5 (Eff 1)
Power diagram Soft-Coagulation
4
0
20
40
60
80
100
120
100 150 200 400 600 1000 2000
Resistance in Ohm
Power in Watt
Soft 1
Soft 2
Soft 3
Soft 4
Soft 5

Table of Contents