25/08/2004 ©
Söring
Service-Manual Arco- and RF-units
1
not valid for A-3000/ -2000
2
not valid for Arco-2000
-55-
3
not valid for MBC600
4
not valid for Arco-1000
Coagulator module
D10 arc recognized
D11 spray-mode (always lights when module starts)
D28 working impulse voltage reached
D29 ignition voltage reached
D56 module overloaded (not in use yet)
P1 Output power (working pulse energy)
P2 Ignition pulse energy
P3 arc detection
Cut module
D14 bipolar mode
D31 arc regulation
D47 power limiting
D50 current limiting
P2 offset power measurement
P3 resonance frequency adjustment bipolar
P4 resonance frequency adjustment monopolar
Handpiece recognition board
D59 interrogation from CPU (must blink periodically)
RF-Output Module (HFO), Piggyback board
P1 adjustment of Offset-power test amplifier (Offset-voltage)
CPU
D1 Idle LED (yellow) shows working of CPU by flickering
D2 Seri LED red (always on)