25/08/2004 ©
Söring
Service-Manual Arco- and RF-units
1
not valid for A-3000/ -2000
2
not valid for Arco-2000
-59-
3
not valid for MBC600
4
not valid for Arco-1000
Cut module
Plug J1
RF-output module
1: RF
2: RF
left at module, to the
right between
cooling
body to RF-
outp.mod.
RF-output module
Plug J1
RF-modules
1: RF (coag module)
2: RF (not
connected)
3: RF (coag module)
4: RF (cut module)
5: RF (cut module)
right bottom
RF-output module
Plug J2
Plug J3
Plug J4
J5
J6
Test resistor
Bipolar connection
PE-connection
Socket Argon
Socket monopolar
2-pole
2-pole
2-pole
1-pole, pin at side
as J5
right over J1
middle top
left bottom
middle left
top left
Handpiece recognition
J1
J2
J3
J8, J9
J7 (power supply
+5V, +15V, -15V)
J13
J14
J16 (I
2
C connection)
Socket monopolar
Socket Argon
Socket bipolar
Foot switch sockets
RF-output module
CPU
Power supply board
CPU (handpiece bus)
CPU (I
2
C bus)
RF-output module
8-pole MICA
8-pole MICA
4-pole MICA
8-pole MICA
5-pole Mascon
1, 2 = GND +5V
3, 4 = -15V GND
4,5 = GND +15V
20-pole MICA
14-pole MICA
8-pole MICA
middle left
bottom left
top left
top right equal val.
top middle
pins from the front
left to right
top right beside J3
top right of J13
top right of J14
CPU board
J1
J2 (status bus)
J3
J5 (mod. control bus)
J6 (HSE bus)
J7 (I
2
C-bus)
J4 (power supply
+5V, +15V, -15V)
Loudspeaker
RF output module
not connected!
Cut- coag module and
high power supply
Handpiece recognition
Handpiece recognition
and front p., gas contr
Power supply board
Handpiece recognition
RF output module.
4-pole RIA
14-pole MICA
14-pole MICA
14-pole MICA
20-pole MICA
14-pole MICA
5-pole Mascon
at the side left
top right
vertical top right
right below J3
top left
right of J6
top middle right
Gas control board
J1 (I
2
C-bus)
J2 (+/- 15V)
J3
J4
J5
CPU
from CPU
Sensors gas pressure
Sensor voltage
+/-15V
Gas pressure switch
Valves
14-pole MICA
3-pole RIA
1 = +15V, 2 = GND
3 = -15V
6-pole RIA
123=+/GND/-Sens. A
456=+/GND/-Sens. B
2-pole RIA
6-pole RIA
1-2 prop. Valve
3-4 valve 2
5-6 valve 1
bottom
between J3 and J4
right of J2
left of J2
left

Table of Contents