Operation
35
Handling microplates
Remove the appropriate plate holder from
the carrier body for loading and unloading
micro plates in the Highplate
®
Rotor 7500
6444.
Before loading, ensure that the rubber bottom is
placed in the cut-outs of the bottom of the plate
holder. Deepwell plates can also be inserted directly
to the carrier without plate holder.
Make sure the rotor is balanced!
The plate carriers are inserted as shown in the illus-
tration below.
Highconic
®
- Rotor 7500 3046
In case longer tubes shall be centrifuged
precluding the complete locking of the
rotor cover, it is allowed to operate the
rotor without cover up to a maximum
speed of 4000 rpm.
Above 4000 rpm the rotor is not allowed to be
operated without cover in any
case. Otherwise the rotor could be
destroyed.