Operation
34
The following examples are intended to illustrate the
different estimates of rotor life for various typical
everyday use:
Profile of use Max. period of use at
5850 rpm 5350 rpm
- frequent use
20 runs / day, 220 days / year
5 years 7 years
- average use
7 runs / day, 220 days / year
14 years 20 years
Always maintain the rotor in the recom-
mended manner. The rotor and accesso-
ries must be clean and inspected regularly:
do not use when showing signs of corro-
sion or cracking.
Highplate
®
Rotor 7500 6444
Do not run the rotor without the rotor cover installed,
Store the Highplate
®
-Rotor with the cover
removed after cleaning. This will enable the
rotor to dry thoroughly.