Troubleshooting
66
Display "OPEN"
appears al-
though lid is
closed.
Will not start.
Lid not properly closed
Open the lid and repeat locking procedure.
Message "Lid"
appears flash-
ingly in the dis-
play.
Drive stops.
Rotor stops without decel-
eration to standstill.
Lid was opened manually during the run.
The instrument stops without deceleration. If you want to continue
the run, you must switch the instrument off and on again.
Message "Lift
Lid“ appears in
the display
Lid does not open auto-
matically.
The lid has not been unlocked after release.
1. Avoid laying objects onto the centrifuge lid.
2. Lift the lid slightly.
E-01
|
E-13
Rotor stops without decel-
eration to standstill.
Instrument cannot be op-
erated.
Internal program error.
Switch the instrument off and on again. If the error persists, contact
KENDRO service.
E-14 Rotor stops with decelera-
tion to standstill.
Instrument cannot be op-
erated.
Overtemperature in the centrifuge tank.
Switch the instrument off and on again. If the error persists, contact
KENDRO service.
Error Symptom Possible causes and corrective measures
!
"
"
"
!
#
$
%
#
&
'
$
(
)
%
&