Troubleshooting
67
E-15
|
E-16
Rotor stops with decelera-
tion to standstill. Instru-
ment cannot be operated.
Temperature measurement error.
Switch the instrument off and on again. If the problem persists, con-
tact KENDRO service.
E-17
Rotor stops with decelera-
tion to standstill.
Instrument cannot be op-
erated.
Max speed for the rotor identification exceeded.
Switch the instrument off and on again. If the problem persists, con-
tact KENDRO service.
E-18 Instrument stops with de-
celeration to standstill after
short starting-up.
Bucket code for this rotor is not defined.
Check, whether the bucket is permitted for the inserted rotor (rotor
program, page 14).
E-19 Instrument stops with de-
celeration to standstill after
short starting-up.
Uncertified rotor.
Check, whether the rotor used is permitted to be operated in this
instrument (rotor program, page 14).
E-20 Instrument does not start
or brakes to standstill.
Rotor cannot be identified.
A) Check whether a certified rotor is inserted.
B) Check whether the rotor is completely fastened onto the motor
shaft. For this release the rotor fastening until you can lift and
depress it easily. Tighten the rotor fastening once again now.
Error Symptom Possible causes and corrective measures