EasyManua.ls Logo

Sotera VISI - To Clean the Visi Mobile Monitoring System Components; To Disinfect the Visi Mobile Monitoring System Components

Default Icon
242 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cleaning and Disinfection
108
To clean the ViSi Mobile Monitoring System components
1. Hand-wash the System components using mild soap or detergent (e.g. Alconox) and water.
Do not appy liqiud to the Cuff Module or Power Pack, instead use a damp cloth.
2. A soft-bristled brush may be used for heavily soiled areas, as needed.
3. Dry thoroughly using a soft cloth or paper towel.
4. Visually examine the reusable components to ensure all soil contaminants have been removed.
5. Repeat the above cleaning process as required.
Note: Use a new Super Sani-Cloth
®
; disinfect according to manufacturer’s recommended
procedure.
To disinfect the ViSi Mobile Monitoring System components
To disinfect the ViSi Mobile Monitoring System components:
1. Disinfect the reusable components by wiping with a Super Sani-Cloth
®
(purple top) or use a basic
wipe moistened with
70% isopropyl alcohol.
2. Dry thoroughly using a soft cloth or paper towel.
Note: Use a new Super Sani-Cloth®; disinfect according to manufacturer’s recommended
procedure.
Do not use bleach, abrasive cleaning agents or organic solvents on any of
the ViSi Mobile Monitoring System components.
Do not autoclave the ViSi Mobile Monitor, its components, or accessories.
Do not use excessive amounts of liquid when cleaning the ViSi Mobile Chest
Sensor CableChest Sensor or the ViSi Mobile Thumb Sensor.

Table of Contents